1: How Walk-in Environmental Test Chambers Work
Walk-in chambers use advanced systems to control temperature, humidity, pressure, and other environmental factors. Key components include:
- Heating and Cooling Systems: Ensure precise temperature control.
- Humidity Generators: Simulate varying moisture levels.
- Control Panels: Allow users to set and monitor conditions.
2: Key Technical Specifications
When evaluating walk-in chambers, consider:
- Temperature Range: Typically from -60°C to +180°C.
- Humidity Range: Usually between 20% and 98% RH.
- Uniformity: Ensures consistent conditions throughout the chamber.
3: Industries That Rely on Walk-in Chambers
- Electronics: Testing devices for thermal and humidity resistance.
- Automotive: Evaluating vehicle components under extreme conditions.
- Aerospace: Simulating high-altitude and space environments.
Product Specifications
Our specialists can help you find the perfect Walk In Environmental Test Chamber.
| Model | DR-H202 |
| System | Balanced Temperature & Humidity Control System |
| Temp. range | -70°C~150°C |
| Humid. range | 20 ~ 98 percent R.H. |
| Temp. & Humid. accuracy | ± 0.3°C; ± 2 percent R.H |
| Temp. & Humid. regularity | ± 1°C; ± 3 percent RH |
| Resolution | 0.1°C, 1 percent, P.I.D control |
| Heating up | -40°C ~ +100°C within 45 min |
| Heating down | 20°C~ -20°Cwithin 40min |
| Interior dimension(W×D×H) | 60×80×85cm |
| Exterior dimension(W×D×H) | 120×135×175cm |
| Interior material | Stainless steel plate(SUS304) |
| Exterior material | Stainless steel plate(SUS304) |
| Insulation | Rigid foam |
| Refrigeration system | Air cooled, Hermetic compressor, single stage refrigeration system. |
| Display | LCD |
| Safety devices | No fuse breaker refrigerator over load relay, Refrigeration High pressure Switch, Over Temp, Protector, Protection relay, Protection fuse, Water Lever protector, Overheat protector, Alarm |
| Accessories | Viewing window, Shelves (freely adjustable 2 pcs.), Cable port(Ø50mm)Chamber lamp. |
| Weight | Approx. 800kg |
| Power | AC380V±10 percent 50/60Hz |
- What is the difference between walk-in and benchtop chambers?
Walk-in chambers are larger and designed for testing full-sized products, while benchtop chambers are smaller and used for smaller samples. - Can walk-in chambers be customized?
Yes, customizable options to meet specific testing needs.